Fume Hood

draftwritten but not verifiedv0.1 · record owner: PI / post-docMott Pro Bench vertical-sash chemical fume hood, model 7321000

Reference record for the FAST lab chemical fume hood. This is a binder and reference document, not a posted sign. The wall sign is Fume Hood.

Provenance is separated on purpose. Values read off the installed unit are not the same as manufacturer catalog values, and neither one is evidence of the hood’s current certified performance. Only the field certification record establishes that.

General-purpose chemical fume hood. FRP-lined. Not a perchloric-acid hood, radioisotope hood, biological safety cabinet, chemical-storage enclosure, or primary containment for explosive reactions.

Field Data
Manufacturer Mott Manufacturing
Manufacturer address 452 Hardy Road, Brantford, Ontario, Canada
Equipment Laboratory chemical fume hood
Product family Mott Pro Bench — Vertical Sash
Model 7321000
Serial number 117290-5-1
Manufacture date December 2023 (12/23)
Nominal hood width 60 in
Liner FRP — fiberglass-reinforced polyester
Sash Vertical rising, stainless-steel cable counterbalance

Electrical data from the nameplate:

  • Primary circuit: 120 V AC
  • Circuit rating: 20 A
  • Frequency: 60 Hz
  • Required circuit breaker or fuse: 20 A
  • Manufacturer warning: disconnect all power supplies before servicing.
  • The hood does not include the fan or exhaust system.
  • The hood may contain more than one live electrical circuit. Consult its wiring diagram before servicing.

Certification markings on the nameplate:

  • CSA c/us
  • CSA file/registration marking: LR 91787
  • UL Classified
  • Classification wording: laboratory hood as to fire, electrical and mechanical hazards only
  • Marking: 68SN — UL 1805

The operating-procedure label on the hood states:

  • DO NOT use perchloric acid in this fume hood.
  • DO NOT use the hood as primary containment for explosive reactions.
  • DO NOT use the hood to store chemicals.
  • Verify the exhaust system and installed monitoring devices are functioning properly.
  • Wear appropriate PPE.
  • Open the sash only as far as necessary.
  • Use the sash as a face and body shield.
  • Keep the sash fully closed when not working.
  • Do not place the head or torso inside the enclosure.
  • Keep the user’s face at least 6 in from the sash.
  • If flammable material is present, remove receptacles and ignition sources from within the hood enclosure.
  • Do not block the bottom airfoil or the baffle openings.
  • Minimize the number and size of objects obstructing airflow.
  • Keep apparatus and associated work at least 6 in behind the sash.
  • Avoid rapid sash, body, arm or hand movements in front of the operating hood.

Everything below this heading is manufacturer catalog and manual data for the model. It describes what the model was designed to do, not what this unit currently does.

Mott’s catalog lists the following standard construction for this family:

  • 60 in nominal width
  • FRP liner
  • Vertical rising sash
  • 6 mm laminated safety glass
  • Stainless-steel sash handle and side runners
  • 30 in viewing height
  • Stainless-steel aerodynamic front airfoil
  • Airfoil notches for routing cords or tubing while allowing sash closure
  • Type 316 stainless-steel round exhaust collar
  • Interior gasketed access panels
  • Removable exterior side panels
  • Corner posts accepting up to five plumbing fittings per post
  • Two UL/CSA-approved duplex receptacles for 120 V service, nominally one per corner post
  • UL/CSA-approved LED light module and switch

The catalog describes the hood as having an automatic compensating upper bypass, with a restricted-bypass option offered for VAV applications. The model number alone therefore does not establish whether the building exhaust serving this hood is operated as CAV or VAV.

  • Width: 60 in
  • Depth: approximately 32-5/8 in
  • Superstructure height: approximately 54-1/4 in
  • Designed for mounting on an approximately 30 in deep countertop
  • Maximum vertical sash opening: 28-1/2 in
  • Exhaust connection, 60 in model: 12 in diameter

These are nominal design dimensions, not an as-built survey of this installation.

Published table for the 60 in model with a 12 in exhaust connection:

Sash opening Average face velocity Exhaust volume Static pressure
28.5 in max 80 fpm 830 CFM 0.15
28.5 in max 100 fpm 1,035 CFM 0.20
18 in 80 fpm 533 CFM 0.05
18 in 100 fpm 667 CFM 0.07

Mott Pro-series specifications express static-pressure loss of this type in inches of water gauge.

These are design reference values. They are not evidence that this hood currently exhausts 830, 1,035, 533 or 667 CFM. The actual certified sash height, face velocity, airflow and alarm setpoint come from the field balancing and certification record.

Mott’s manual states its fume hoods are intended to provide containment at roughly 80–120 fpm face velocity, with some models capable of lower operation, and that actual exhaust requirements depend on sash position.

Model 7321000 is the FRP-lined version. Mott describes FRP as a general-purpose liner resistant to a broad range of reagents, but states that FRP-lined hoods are not suitable for perchloric-acid or radioisotope work. Mott also states chemical fume hoods should not be used for biological work, chemical storage, spray painting, or processes incompatible with the liner material.

Mott’s manual specifies 12 AWG minimum supply wiring for a 20 A circuit, permanent grounding, and a nearby accessible dedicated disconnect or breaker. Hoods not factory prewired by Mott are to be field-wired to applicable electrical and local codes.

  • Locate apparatus as far toward the rear as practical, and never closer than 6 in (15 cm) from the sash plane.
  • Elevate large apparatus approximately 2 in (5 cm) above the work surface where practical, to allow airflow underneath.
  • Do not overload the hood. Mott gives a rule of thumb of no more than 50% of the available work surface covered by equipment or materials.
  • Avoid rapid movements at the hood face.
  • Avoid pedestrian traffic immediately in front of an operating hood.
  • Do not adjust baffles. This is work for qualified airflow or safety personnel.

Before each use:

  • Verify exhaust operation.
  • Verify the airflow alarm/monitor is operating.
  • Check that airflow openings are unobstructed.
  • Use the established certified sash working height.

At least annually:

  • Containment testing, which Mott recommends before first use and at least annually after.
  • Inspect internal sash pulleys and cables for corrosion, fraying or deterioration.
  • Inspect the cable attachment to the sash handle.
  • Lubricate cables with lithium grease yearly or as required.
  • Inspect more frequently under intensive use or in corrosive environments.

If the low-flow alarm activates:

  • Close the sash immediately.
  • Step away.
  • If airflow does not recover and the alarm does not clear, evacuate the affected area and contact facilities. An exhaust-system failure may exist.

For the hood’s electrical equipment:

  • Indoor use only
  • Pollution Degree 2
  • Installation Category II
  • Maximum altitude: 3,000 m ASL
  • Operating ambient temperature: 5–40 °C
  • Maximum RH: 80% through 31 °C, decreasing linearly to 50% RH at 40 °C

None of the following can be inferred from the model number. Read them off the hood, the facilities stickers, or the certification record. These describe the actual current performance of this hood and matter more operationally than most of the catalog data above.

  • Room and hood asset ID
  • Last certification date
  • Next certification due date
  • Certified sash operating height
  • Measured average face velocity
  • Measured exhaust CFM, if reported
  • Airflow alarm model
  • Alarm setpoint
  • Building exhaust/fan identifier
  • CAV or VAV configuration
  • Exhaust fan, belt and service information
  • Installed plumbing services
  • Installed work-surface material
  • Emergency electrical disconnect location
  • Facilities and service contact
  • The manufacturer data in this record was compiled from an assistant-generated summary of the Mott documents listed above, not read from those documents directly. Confirm the configuration, dimensional, exhaust-performance, wiring, setup, inspection and environmental figures against the manual and catalog PDFs before the record is relied on.
  • Confirm which Mott document is the source for the 80–120 fpm containment range, the 12 AWG supply wiring figure, the annual cable inspection and lubrication intervals, and the rated environmental conditions. The summary attributed these to “Mott’s manual” without a specific citation.
  • Confirm the nameplate and operating-label transcriptions against the physical unit.