Skip to content

Well Plate Pipette Guide

A parametric, 3D-printable pipette guide for laboratory well plates, written in OpenSCAD. It sits over a standard plate and steers the pipette tip into the centre of each well, through a tapered hole and a short cone that reaches into the well. That keeps the tip off the well wall and the sample, which matters for delicate work such as organoid culture.

By Cameron Brooks and Hadden Christ, 2026. The guide has been printed and used with 24- and 96-well plates.

pipette_guide.scad draws three parts around one plate definition:

  • Stand: a frame that sits around the well plate, with arched cut-outs in all four sides.
  • Guide: the plate the pipette goes through, which sits on the stand. Each well gets a tapered hole, wider at the top, inside a cone that enters the well. Under the guide, a lip with small snap bumps locates it in the stand. Lift tabs on the ends make it easy to take off, and row letters and column numbers are engraved in the top.
  • Base (optional): a tray the stand nests in, with a hole under each well.

With guide_interstitial_protrusions = true (the default), the guide also has solid cones, with no hole, between each group of four wells. They press on the plate to hold it down.

The plate definitions are in well_plates.scad. plate_selection picks one:

plate_selection Plate Wells Pitch Well diameter Default guide fits
0 24 CytoView MEA 4 x 6 18.00 mm 15.00 mm yes
1 96 flat short 8 x 12 9.00 mm 6.86 mm yes
2 96 round medium 8 x 12 9.00 mm 6.86 mm yes
3 384 flat short 16 x 24 4.50 mm 3.63 mm no
4 1536 flat short 32 x 48 2.25 mm 1.50 mm (square) no

The default holes and cones are sized for the 24- and 96-well plates. On the 384- and 1536-well plates, the cones (5.97 mm across where they enter a well) are wider than the wells, and the render stops with a message naming the parameter to change. A guide for those plates needs smaller holes and cones, which nobody has designed or printed yet.

A plate definition is one array. The comment block in well_plates.scad lists each field: rows, columns, pitch, the A1 offsets, well diameter, depth and shape, bottom radius and flange heights. The grid and the outer dimensions follow the ANSI/SLAS microplate standards, cited in references/. The well dimensions come from the plate maker’s datasheet. To add a plate, add an array in the same format to well_plates.scad and to the well_plates list, and add it to the plate_selection options in pipette_guide.scad.

The main ones, at the top of pipette_guide.scad, with their defaults:

Parameter Default What it sets
render_part "assembly" "assembly", "stand", "guide" or "base"
render_well_plate true draws a transparent reference plate
view_cross_section false cuts the view in half to show the holes
plate_selection 0 the plate, from the table above
tip_diameter 3.6 mm the top of each guide hole
bottom_hole_diameter 2.6 mm the bottom of each guide hole
protrusion_extension 3 mm how far the cones reach into the wells
protrusion_top_radius tip_diameter the cone’s outside radius at the guide
protrusion_taper_angle 5.5 deg how much the cone narrows towards its tip
guide_interstitial_protrusions true the solid cones between wells
fit_clearance 0.2 mm clearance between the plate, stand and guide
label_depth 0.5 mm how deep the labels are engraved

To print, set render_part to one part at a time and turn render_well_plate off.

tip_checker.scad is a small test plate for finding the hole size that suits your tips before printing a whole guide. It has an 11 x 11 grid of tapered holes: the top diameter steps from 6.0 to 8.0 mm across the columns, and the bottom diameter from 4.0 to 6.0 mm down the rows, both by 0.2 mm. Both are engraved on the plate. These are the same quantities as the guide’s tip_diameter and bottom_hole_diameter, so the numbers of the hole that fits go straight into the guide.

The checker’s range does not include the guide’s defaults (3.6 and 2.6 mm). To test around other sizes, edit top_diameters and bottom_diameters at the top of the file.

  1. Print the tip checker and try your pipette tips in it.
  2. Set tip_diameter and bottom_hole_diameter to the hole that fits best.
  3. Set plate_selection to your plate.
  4. Render and export the stand, the guide and, if you want it, the base, one at a time.
  5. Print them, and try the guide on an empty plate before using it with samples. It should sit flat, and the tip should pass through without force.

Render with the Manifold backend (openscad --backend=Manifold, or the Manifold option in OpenSCAD’s preferences). Each part of the 24- and 96-well guides rendered in 2 s or less with OpenSCAD 2026.10.15.

File Contents
pipette_guide.scad the stand, guide and base, and their parameters
well_plates.scad the plate definitions
well_plate.scad a model of a standard plate, drawn from a definition
tip_checker.scad the hole-size test plate
utils.scad cross_section(), used for the cross-section view
references/ the ANSI/SLAS standards the plate geometry follows
CHANGELOG.md changes by version

GNU Affero General Public License v3.0. See LICENSE.