Skip to content

Syringe Actuator: Design

Status: every syringe builds and passes just test and just qa. N20 units of the per-syringe design ran as designed, one for 2 hours with no wear (2026-10-07); the shared N20 body is not yet printed; the 10 and 30 mL are built from estimates and not yet printed. The syringe figures below were measured by hand unless marked as estimates; the motor figures are measured or come from the drawings named beside them.

Printed parts that connect a DC gearmotor to a luer syringe so the plunger strokes back and forth continuously: an N20 for the 1, 3 and 5 mL, a 25 mm GA25 for the 10, 20 and 30 mL. The motor runs in one direction straight off a supply; each revolution is one full pull and one full push.

Not in scope: what the syringe is connected to, anything that powers or controls the motor, and the adapters a user mounts the body with. The body has two mounting patterns for them.

An in-line crank-slider. The motor’s shaft stands vertical on the line of the syringe’s axis, so the crank, the rod and the plunger all move in one horizontal plane. Figures for one syringe are the 3 mL’s; part="report" gives any syringe’s.

  • Frame. A 5 mm plate as wide as the widest barrel flanges in its body’s group (27.3 mm for the N20 syringes, the 5 mL’s), which drop into a slot at one end; the slot’s walls take both the push and the pull. The barrel snaps into a ring 15 mm in front of the flange’s block, past two arms 1.6 mm thick that stand free from 45 deg below the axis, so they bend over a long arc. The opening is ring_grip (0.5 mm) a side narrower than the barrel, with a 45 deg lead-in at each tip. Two rails, 10 mm tall, run from the slot’s back wall either side of the clip’s travel. The motor hangs under the plate’s other end, its gearbox 2 mm up into a pocket, located by its bushing and held by two screws counterbored under the crank’s hub. The screws lie along the stroke, so they resist the rod’s push and pull: an N20 stands with its 12 mm side along x.
  • Cradle. A sleeve under the plate that grips the gearbox for its whole length below the plate, 10.5 mm on an N20 and 24.8 on a GA25. The rod’s force acts about 15 mm above the gearbox’s face and reverses every stroke; with only the 2 mm pocket and the two screws to stop it, the motor rocks, and the screws work loose. Two set screws through trapped nuts push the gearbox against the far walls, so it has no play to rock in: on an N20, one through the front wall and one through the -y wall; on a GA25, two through the front, 30 deg either side of -x. Two M3 x 8 screws hold it to the plate from below, through flanges at its front and back, threaded 4 mm into the plate. The plate widens where the cradle needs it to pass between the body’s rails. It prints top face down.
  • Crank. It presses onto the motor’s D-shaft and carries the pin at a throw of half the syringe’s stroke less a 1 mm margin at each end: 27.6 mm for the 3 mL. Its D-bore is blind, so it stops on the shaft’s tip, 1 mm over the plate. An M3 x 6 grub screw (ISO 4026, cup point) runs back through a lobe behind the hub onto the shaft’s flat, which faces it, through an M3 nut dropped into a slot from the top: steel on steel carries the torque and holds the crank down.
  • Rod. At least three throws long, 90 mm at the shortest, from the crank pin to the clip, lying on both parts’ top faces. Each end turns on a printed boss 0.3 mm taller than the rod; an M3 x 10 screw and washer clamp the boss, not the rod, so the joint turns freely with the screw tight.
  • Plunger clip. The push-button drops into a cup and snaps past two lips (grip in the Customizer), so the syringe stays in whichever way up the unit is; the plunger passes a slot in the cup’s front wall. A sled under the clip runs 20 mm past the rod’s pin between the frame’s rails, 0.3 mm a side, 2 mm below the rod.
  • Body. A shell fitted round what the frame carries, symmetric about the axis’s vertical plane, square at the top and chamfered at the bottom as far as its contents allow. layout() sizes each section from the extremes of what the frame carries over a whole turn.
    • Shared. The N20 syringes (1, 3 and both 5 mL) share one body, lid and cap, sized for the largest of what each needs, measured from the motor’s shaft. Their axis height and plate width are the group’s largest, and each frame’s plate runs forward to the shared front wall, so swapping syringes reprints only the frame, clip, crank and rod. The group is the fourth column of syringes; the GA25 syringes still get a body each.
    • Shape. Narrow and shallow at the front, where there is only the syringe, the plate and the clip; full width and the motor’s depth where the crank turns; a taper between, which layout() checks against the rod’s swing and keeps within 45 deg of the body’s axis.
    • Frame and cap. The frame slides in from the back on rails that take the plate’s dovetailed edges, up to the front wall. A cap screwed to the back end (2 x M3 x 10) holds it there with a post in the same rails, and lets the motor’s wires out.
    • Window and lid. The syringe goes in and out through a window in the roof at any clip position, the barrel through a slot in the front wall; the window ends in a 45 deg point. A lid slides on rails along the roof between a lip on the front wall and the cap; a bump under its back end clicks into a notch when it is shut and runs in a groove the rest of its travel. Over the wider back the rails run out to the sides, filling the roof’s shoulders flush with their tops, so they cannot snap off and leave no inside corner to clean.
    • Mounting. Two patterns of four M3 clearance holes, each sealed by a 0.6 mm knockout skin at the outer face until a user drills or punches the ones they use; the screws’ heads sit inside. One is through the front wall beside the barrel’s slot, 10 up and the narrowest multiple of 10 across whose heads clear the frame’s snap ring, for the widest ring in the body’s group: 30 for the N20 syringes and the 10 mL, 40 for the 20 and 30 mL, centred 15 mm over the plate’s underside; the other through the roof’s shoulders, 60 across by 20 along, centred 15 mm in front of the back end, behind the crank’s sweep. Another OpenSCAD project draws its adapter from the layout with use <cad/main.scad>: actuator(syringe, motor) gives it, front_x() and the body’s front_face() the front wall’s outer face, the body’s mount_front and mount_top the holes, and tip_x() the syringe’s front end, from its nozzle_l.
    • Printing. The body prints standing on its front wall with no supports: 242 mm tall for the N20 syringes. The front wall is what the frame stops against, so it prints flat on the bed rather than bridged. Standing that way, nothing overhangs past 45 deg: the walls lean in at most that, the window ends in a 45 deg point, and the cap’s bosses taper into the wall; the four roof mounting holes print as small horizontal holes. Turn on the slicer’s elephant-foot compensation, and use a brim, as the body stands tall on a small footprint. No edge on the body, the lid or the plate comes to a knife edge: the dovetails and the lid’s bevels stand on short lands.
    • Assembly. Front mounting screws, if any; the frame into the body from the back; the lid onto its rails from the back; then the cap. Top mounting screws go in after the frame, or before it if the frame slides in with its crank along the axis, as a crank turned across carries its pin screws into them.

The motor never reverses. At a = 180 deg the plunger is in, stopping 1 mm short of the barrel’s end, and at 0 deg it is out, stopping 1 mm short of the plunger’s stop. The margin at each end takes up error in the measured exposed length and stroke, and in the print.

Side load. The rod’s angle phi puts a side load F * tan(phi) on the clip, where F is the force along the stroke. The rod is at least three throws long, so phi peaks under 19.5 deg: asin(27.6/90) = 17.9 deg on the 3 mL. The rails take that load; the plunger cannot carry a side load in compression.

N20, for the 1, 3 and 5 mL. A 12 V N20 at about 30 rpm, a 1000:1-class gearbox: Pololu #3046 or a generic “N20 12V 30RPM”. Pololu gives #3046 35 rpm free-running and a 10 kg.cm extrapolated stall, and recommends keeping instantaneous torque under 2.5 kg.cm on this gearbox (https://www.pololu.com/category/60/micro-metal-gearmotors). 35 rpm gives a 1.7 s cycle and, on the 3 mL, a peak plunger speed of about 101 mm/s. A lower supply voltage slows it roughly in proportion.

GA25, for the 10, 20 and 30 mL. A 20 mL plunger, pushed by hand on a kitchen scale with the tip open (2026-10-02, rough), slides at about 250 g (2.5 N), and about 500 g (4.9 N) pushed hard and fast: 1.0 and 1.9 kg.cm on the 20 mL’s 39 mm throw. The GA25YN370 at 12 V and 30 rpm gives a 12 kgf.cm stall by its seller, and Pololu recommends at most 4 kg.cm continuous on its 25 mm metal gearboxes. Rejected: the yellow plastic “TT” gearmotor (about 0.8 kg.cm stall at 6 V: Adafruit’s datasheet for product 3777), a 20 mm “turbo worm” motor whose listing’s 8 kg.cm a buyer measured at under 1, and a 37 mm JGB37-520, with more margin than the load needs at about three times the cost.

Which motor. Friction differs little between sizes, but vacuum or back-pressure acts on the bore’s area, which grows fast: the 10 mL’s is nearly three times the 3 mL’s. Peak crank torque is about the plunger’s force times the throw; estimated from 3 N of seal friction up to 10 mL and 5 N from 20 mL (assumed, bar the 20 mL’s scale test) and BD’s published bores:

Syringe Open line Against full vacuum (1.013 bar)
1 mL 1.0 kg.cm 1.5 kg.cm
3 mL 0.8 kg.cm 2.5 kg.cm
5 mL, the first 0.8 3.8
5 mL, the second 0.7 3.4
10 mL 1.1 6.9
20 mL 2.0 13.6
30 mL 2.3 19.4

On an open line the N20 drives up to the 10 mL; against vacuum or pressure it suits only the 1 mL, and the 3 mL right at its limit. The 10 mL’s 6.9 kg.cm against vacuum is under the GA25’s 12 kg.cm stall but over the 4 kg.cm continuous limit the report holds it to, and a blocked 20 or 30 mL stalls either. Positive pressure has no ceiling, unlike vacuum. So motor in the Customizer is auto, the N20 up to the 5 mL and the GA25 from the 10 mL, or either by name, and the report gives the estimated torque for the syringe and motor chosen beside that motor’s limit; it assumes 3 N of friction throughout, so for the 20 and 30 mL it gives less than the table.

The 3 mL measured 2026-09-29 (its barrel length and stroke 2026-10-02), the 1, 5 and 20 mL 2026-10-02, a second 5 mL before 2026-11-06. They are not BD’s: the 3 and 5 mL carry no maker’s mark (BD prints its logo at the scale’s end), and another maker’s syringe of the same volume can differ by most of a millimetre, so a syringe from elsewhere is measured and added as its own entry (measuring.md).

Feature 1 mL 3 mL 5 mL 20 mL
Barrel OD 9.45 mm 10.0 mm 13.6 mm 20.9 mm
Barrel, flange front face to the shoulder at the nozzle’s root 75.0 mm 68.0 mm 63.5 mm 91.75 mm
Barrel flanges 24.6 x 12.6 mm, 2.0 mm thick 24.5 x 12.6 mm, 2.0 mm thick 26.9 x 16.0 mm, 1.4 mm thick 36.3 x 24.2 mm, 1.6 mm thick
Push-button dia 10.2 mm, 1.8 mm thick dia 12.6 mm, 1.6 mm thick dia 16.6 mm, 1.8 mm thick dia 20.9 mm, 1.4 mm thick
Working stroke 64.25 mm 57.25 mm 52.5 mm 80.0 mm
Plunger exposed, stopper bottomed (flange rear face to push-button front face) 10.75 mm 10.75 mm 14.3 mm 11.75 mm
Plunger, widest across (each narrows a little towards the push-button) 3.8 mm, a solid rod 7.5 mm, ribs about 1 mm thick; the last 9 mm before the push-button tapers to 6.6 mm 11.5 mm 18.3 mm
Nozzle, the barrel’s shoulder to the syringe’s front end not measured not measured 10.2 mm, derived from the assembled unit, not measured not measured

A second generic 5 mL (5ml-gnr2) matches the table’s across the axis and is shorter: stroke 47.3, exposed 13.6, barrel 56.8 and nozzle 10.0 mm, measured. Its row’s crank_as points at the first’s row (5ml-gnr1), so layout() gives it the rod, 93.3 mm, that keeps the crank where the first has it: a unit printed for the first takes it with only a new crank and rod, and the frame, clip, body, lid and cap carry over. Its nozzle’s tip then sits 6.9 mm further back from the body’s front face (32.7 mm past it against 39.6), so an adapter drawn to the first’s tip_x() does not reach it. The table’s 5 mL column is the first.

BD Plastipak luer-lock syringes, estimated and not measured: from BD’s photo of its range, scaled against the 5 and 20 mL above, with outside diameters from published figures (Fisher’s 16.21 mm for the 10 mL, 23.77 and 29.5 mm for the 30 and 50 mL) and exposed lengths from a Terumo pump manual’s flange-to-plunger-end figures less the push-button. Good to about a millimetre; measure one in hand and replace them. The 10 and 30 mL build; the 50 mL is not built.

Feature 10 mL 30 mL 50 mL
Barrel OD 16.2 mm 23.8 mm 29.4 mm
Barrel, flange front face to the shoulder at the nozzle’s root 82.5 mm 104.5 mm 130.0 mm
Barrel flanges 30.0 x 19.0 mm, 1.5 mm thick 44.0 x 27.5 mm, 1.8 mm thick 53.0 x 33.2 mm, 2.2 mm thick
Push-button dia 18.0 mm, 1.6 mm thick dia 26.0 mm, 1.8 mm thick dia 31.0 mm, 2.0 mm thick
Working stroke 71.0 mm 92.5 mm 117.0 mm
Plunger exposed, stopper bottomed 11.5 mm 14.5 mm 15.0 mm
Plunger, widest across 13.7 mm 20.8 mm 25.7 mm

The parts drive 2 mm less than the working stroke, stopping 1 mm short of each end. The barrel must be longer than the stroke plus that margin, or the stroke pulls the stopper out; layout() asserts it.

The part names follow ISO 7886-1: barrel, barrel flanges (finger grips), plunger, push-button (thumb rest), plunger stopper (the rubber piston) and nozzle (the luer tip).

From Pololu’s micro metal gearmotor dimension drawing (https://www.pololu.com/file/0J949/micro-metal-gearmotor-dimensions.pdf, 2024-04-03), except the shaft’s length, measured on the motor in hand (2026-09-29):

Feature Size
Can 12.0 wide x 10.0 across flats x 15 long; end cap 1.2, then a dia 5 boss
Gearbox 12.0 x 10.0 x 9 long (12.5 at 1000:1)
Shaft dia 3.0, 2.5 across the flat; 9 long from the gearbox face (measured), D-profile over its whole length
Bushing dia 4.0, 0.7 proud
Mounting 2 x M1.6 tapped, 9.0 apart, either side of the shaft

Held by two M1.6 x 3 JIS flat heads (McMaster 91305A002), which sit in counterbores sized for an ISO 7045 pan head, their cone on the clearance hole’s rim, 1.4 mm into the gearbox, and by the cradle under the plate.

HERCHR, Amazon.ca B0BX6B91L9; the same part is sold by aqxreight (B0BSC9CK13) and as the JGA25-370 and 25GA-370. Measured on the motor in hand, 2026-10-03; it agrees with Seeed’s JGA25-370 drawing (https://media.digikey.com/pdf/Data%20Sheets/Seeed%20Technology/114090046_Web.pdf) bar the gearbox’s length, longer at this ratio. Ratings from the HERCHR listing.

Feature Size
Gearbox dia 25 x 26.8
Can dia 24.3 x 30.8, and 1.75 of end boss and terminals past it
Shaft dia 4.0, centred, 12.5 from the gearbox face (10 past the boss); 3.5 across the flat, the flat 8 long from the tip
Boss dia 7 x 2.5
Mounting 2 x M3 tapped, 17 apart across the shaft, at most 3.5 deep (Seeed’s drawing; not measured). The face’s other two screws hold the gearbox together: leave them in
Rating (listing) 12 V, 30 rpm no load; 25 rpm at 770 gf.cm; stall 12 kgf.cm at up to 2.1 A

Its gearbox goes 2 mm up into a round pocket, held by two M3 x 4 button heads (ISO 7380), 2.85 mm into its holes: the plate over the pocket is 3 mm, too thin for socket heads. The crank’s bore is D-shaped only as far as the shaft’s flat runs, round below. The motor hangs 57 mm under the plate, so its bodies are deep: the 5, 10, 20 and 30 mL’s stand 223, 256, 277 and 313 mm tall as they print.

No usable OpenSCAD model of either motor was found (NopSCADlib has none), so cad/purchased/motor.scad draws them, simplified, from the figures above.

  • The rod’s side load needs a guide. On the first print the clip had no rails; on the push stroke the side load bowed the plunger sideways and the clip turned with it (video, 2026-09-30).
  • A short snap arm breaks. The first snap rings, 2 mm thick and free only above the axis with an opening 80% of the barrel’s width, snapped in PLA on every size as the barrel went in, worse as it grew and squashed less (reported 2026-10-05); a PETG one with its tips cut back held. The ring above works in PLA and in PETG (2026-10-07). ring_test prints the ring alone on a stub of plate, to try the snap on a syringe in a few minutes.
  • A printed D-bore does not hold a shaft. Keyed only by the flat’s half millimetre, a PETG crank’s bore wore round and rode up an N20’s shaft within an hour, as the load reverses every stroke (reported 2026-10-05). With the grub screw, cranks printed in PLA and in PETG held, and a 2-hour run left the bore unworn (2026-10-07).
  • A crank’s set screw, tightened hard, tips the crank. On the unit below, an overtightened grub screw tipped the crank on its shaft and stalled the stroke (2026-11-06 to 11-07).
  • Two screws do not hold a motor against a reversing load. A 5 mL unit (5ml-gnr2) on an N20 drove a turbidity column’s sampler, drawing a full 5 mL of algae culture from an unpressurized, sparged column through a 3 mm riser: one 18.5 s burst every 15 minutes, 1 s at full throttle then 17.5 s at half. From its first burst, 2026-11-06 22:46, to the last that moved liquid, 2026-11-15 08:19, it ran 8.40 days, 863 bursts, 4.43 h with the motor on. Its motor’s screws worked loose until the motor rocked, and the crank’s D-bore cammed out on the shaft; threadlocker only delayed it. Each burst’s time is from its own 50 Hz log rows, in turbidity-sensor. The screws then lay across the stroke, on the line the motor tipped about, so only the 2 mm pocket held it. The motor now stands with its screws along the stroke, and the cradle grips the gearbox; the 5 mL also builds on the GA25 (5ml-gnr1-ga25, 5ml-gnr2-ga25) while both are tested.
  1. Measure the 10 and 30 mL syringes in hand and replace their estimates.
  2. Measure the GA25’s mounting holes’ depth. Its screws are sized to Seeed’s 3.5 mm.
  3. Measure the N20’s end boss and terminals past its can; the body allows 3 mm, nominal.
  4. Measure nozzle_l, the barrel’s shoulder to the syringe’s front end, on the 1, 3, 10, 20 and 30 mL, where tip_x() stops the build without it, and on the first 5 mL, whose figure is derived.
  5. Measure the plan shape of the barrel flanges. The slot is open at the sides and sized only by the flanges’ width, height and thickness.